OPTIC-4 Multi Mode Inlet

GC에 장착되어 다기능의 시료주입과 시료전처리 option이 가능한 Muti-Mode inlet

구분
실험실용
제조사
GLSCIENCE-네덜란드 이미지 (GLSCIENCE-네덜란드)
제품관련자료

MultiMode Inlet for any Gas Chromatograph Optic 4

1. OPTIC-4 is the ultimate among high performance inlet systems for gas chromatography, types Designed to be installed easily onto virtually any make or model of gas chromatograph.

2. OPTIC-4 offers the widest range of injection modes for a vast array of sample

3. The patented* design of the injector body helps to have the best transfer of the compounds from the inlet onto the analytical column. This can be seen both for volatile and non-volatile compounds.OPTIC-4 Multi Mode Inlet can cool very fast because of the low thermal mass. The GC inlet system can work from cryogenic temperatures (-150°C) up to very high temperatures (600°C) with a ramp rate up till 60°C/sec.Depending on solvent type, liner type and injection speed it is possible to inject from 0.1μl till 500μl in the OPTIC inlet. With the optional inlet peltier cooler it is possible to cool faster and reach initial temperatures of +7°C.

4. Increase capacity of your GC system with MultiMode OPTIC
(1) Cold injection
(2) Large volume injections (sorbent vent)
(3) Thermal Desorption
(4) Pyrolysis (solid & liquid)
(5) High temperature injections
(6) On column injections
(7) ColdTrap / Cryogenic Trap / Cold Trap Gas Chromatography
(8) In-injector derivatization
(9) Compatablity with PAL RTC

5. Cold Trap Option for OPTIC4
(1) Operating using direct LN2 or CO2
(2) Temperature range: ?-150 °C to 600 °C
(3) Temperature stabilaty at low temperature: ?±3 °C
(4) Heating ramp rate: 1 to 60 °C/sec
(5) Cool down time:2-3 minutes
(6) Split/Splitless

OPTIC.jpg

6. Liner Exchanger with CDC(Capping, De-capping) Option
(1) Normal liquid injection without different configuration
(2) No transport adopter, just normal liners/sample tubes
(3) Little if any sample preparation is required
(4) Automated with a PAL System
(5) Selective desorption of components without pyrolyzing of the matrix
(6) Low consumable costs
(7) For any type GC system
(8) Normal septum purge (no septum bleeding and carry over)
(9) Uses standard 11mm septa and is compatible with the Merlin Microseal
(10) Tray capacity PAL RTC 54 and 40 for capped liners